Development of the quality assurance/quality control procedures for a neutron interrogation system

Obhođaš, Jasmina; Sudac, Davorin; Valković, Vladivoj (2016) Development of the quality assurance/quality control procedures for a neutron interrogation system. IEEE Transactions on Nuclear Science, 63 (3). pp. 1544-1550. ISSN 0018-9499

[img]
Preview
PDF - Accepted Version - article
Download (3MB) | Preview

Abstract

In order to perform Quality Assurance/Quality Control (QA/QC) procedures for a system dedicated to the neutron interrogation of objects for the presence of threat materials one needs to perform measurements of reference materials (RM) i.e. simulants having the same (or similar) atomic ratios as real materials. It is well known that explosives, drugs, and various other benign materials, contain chemical elements such as hydrogen, oxygen, carbon and nitrogen in distinctly different quantities. For example, a high carbon-to-oxygen ratio (C/O) is characteristic of drugs. Explosives can be differentiated by measurement of both (C/O) and nitrogen-to-oxygen (N/O) ratios. The C/N ratio of the chemical warfare agents, coupled with the measurement of elements such as fluorine and phosphorus, clearly differentiate them from the conventional explosives. Here we present the RM preparation, calibration procedure and correlations attained between theoretical values and experimentally obtained results in laboratory conditions for C/O and N/C ratios of prepared hexogen (RDX), TNT, DLM2, TATP, cocaine, heroin, yperite, tetranitromethane, peroxide methylethylketone, nitromethane and ethyleneglycol dinitrate simulants. We have shown that analyses of the gamma ray spectra by using simple unfolding model developed for this purpose gave a nice agreement with the chemical formula of created simulants, thus the calibration quality was successfully tested.

Item Type: Article
Uncontrolled Keywords: Calibration procedures; QA/QC; container inspection; neutron interrogation; threat material analysis; threat material simulants
Subjects: NATURAL SCIENCES > Physics
Divisions: Division of Experimental Physics
Projects:
Project titleProject leaderProject codeProject type
Razvoj i primjene nuklearnih analitičkih metoda-Jasmina Obhođaš098-0362975-2881MZOS
RAZVOJ METODA KONTROLE PROMETA OPASNOG MATERIJALA-Dario Matika098-0362975-2891MZOS
Depositing User: Jasmina Obhođaš
Date Deposited: 25 Oct 2016 14:04
URI: http://fulir.irb.hr/id/eprint/3088
DOI: 10.1109/TNS.2016.2521900

Actions (login required)

View Item View Item

Downloads

Downloads per month over past year

Contrast
Increase Font
Decrease Font
Dyslexic Font
Accessibility